01 · PEM Hydrogen Generators
HG-250
250 ml/min for a single instrument
The smallest of the three capacities: 250 ml/min at full output and 175 ml/min for continuous consumption. Water is electrolysed across the proton-exchange membrane as the gas is consumed, so the only hydrogen in the instrument is the small volume in the filter train.
250 ml/min
Max. generation
175 ml/min
Continuous duty
99.9995 %
Purity

Overview
What this model does well
Gas made as it is drawn
Nothing is stored under pressure beyond the small volume held in the filters — there is no cylinder inventory in the lab.
Three-stage drying train
Water trap, silica gel and a 5 Å molecular sieve bring the gas to 99.9995 % purity at the outlet.
Deionised water, no electrolyte
Below 0.5 µS/cm, with no KOH or NaOH added. Water quality is shown on the LCD in set mode.
Alarms that stop generation
Water empty, water over and pressure over each stop the cell: the Alarm LED blinks, the buzzer sounds every 10 s and the message appears on LCD line 4.
Specifications
HG-250 — full specification
Compare with other models →Generation
- Maximum H₂ generation
- 250 ml/min
- Continuous duty (70 % of max.)For continuous consumption, draw no more than 70 % of the model's maximum capacity.
- 175 ml/min
- H₂ purityReached after the three-stage drying train.
- up to 99.9995 %
- Cell technology
- PEM electrolysis of pure water
- Hydrogen storage
- None — only the volume held in the filters
- By-product
- O₂, vented
Pressure
- Outlet pressure
- Set on the keypad
- Over-pressure cut-outAbove 90 PSI generation stops and the pressure over alarm is raised.
- 90 PSI
Feed water
- Water quality
- Deionised, < 0.5 µS/cm (> 2 MΩ·cm)
- ElectrolyteAdding alkali destroys the membrane and voids the warranty.
- None — never add KOH or NaOH
- Tank level sensing
- 3 levels — low, medium, full
- Water change
- Empty, rinse and refill every 30 days
Drying train
- Stages
- Water trap + silica gel + 5 Å molecular sieve
- 1 · Water trap
- Automatic liquid knock-out, solenoid drain — no service medium
- 2 · Silica gelRegenerate in an oven at 120 °C for 60–120 min.
- Blue indicating gel; pink at 60 % saturation
- 3 · Molecular sieveRegenerate in an oven at 300 °C for 120–240 min.
- 5 Å, dark yellow; no colour change — service monthly
Control & interface
- Display
- 4-line LCD, normal and set mode
- Keypad
- 3 keys — ▲, ▼, F (pressure / overflow)
- Status LEDs
- Power, Generate, Alarm
- Water-level LEDs
- Full, Med, Low
- Readings in set mode
- Operating hours, water quality, total H₂ generated
- AlarmsAlarm LED blinks, buzzer sounds every 10 s, message on LCD line 4.
- Water empty, water over, pressure over, H₂ leak
- Data interface
- RS-232
Installation & supply
- Mains supply
- 220 V AC, 50 Hz
- Power / fuseRating as supplied for the HG-250. Confirm the supply rating for HG-500 and HG-1000 at order.
- 150 W · 5 A glass fuse
- Rear connections
- H₂ outlet, water drain, RS-232, cooling fan
- Siting
- Level bench, fan and rear panel clear, drain accessible
- Service interval
- Water change + tank clean every 30 days
Ordering
- Order code
- HG-250
- Overflow control (HG-OFC)
- Option — kit with three-port N₂ purge valve
- Filter refill set (HG-FLT)
- Silica gel and 5 Å molecular sieve
Values taken from HC-DS-PEMHG Rev 1. The configuration is confirmed in your offer.
Documentation
Datasheet & documents
Primary document
PEM-HG datasheet
PDF · EN · 3.2 MB · HC-DS-PEMHG Rev 1
Model matrix for HG-250, HG-500 and HG-1000, gas path and drying train, control panel and alarms, connections, options and order codes.
Download PDF- PDFDownload ↓
PEM-HG datasheet
EN · 3.2 MB · HC-DS-PEMHG Rev 1
- PDFDownload ↓
PEM-HG operating manual
EN · 2.8 MB · HC-OM-PEMHG Rev 1
The PEM-HG datasheet and operating manual are published in English and can be downloaded directly. German editions are in preparation — ask us and we will send them as soon as they are released. Go to the downloads centre →